Hind Vactech Scientific Pvt.Ltd.

+91 99588 22058 & +91 97160 14312

Our Product Range

Vactech Vacuum Pumps

Laboratory Cooling Equipments

Laboratory Heating Equipment

Laboratory Testing Equipment

Vacuum Gauges

Lubricating Oils

Fittings & Flanges

DIGITAL VACUUM & TEMPERATURE CONTROLLER 

MODEL: DTVC-001

The UVS Model DTVC series Vacuum & Temperature Gauge Controller displays vacuum pressure as well as temperature in one module, the vacuum is measured from a S-PR-GH sensor tube. It displays vacuum measurements based on thermal conductivity. The input for the temperature channel is from the RTD thermocouple. The Model DTVC-001 controller controls relay with independent set point.
***The optional RS485 along with data acquisition software can be obtained on request

  • Measurement range 0.001 to 1000 mbar for vacuum and -100 to +800 Deg C for Temp
  • Space saving rugged design DIN console
  • Mounts in any orientation
  • brass body measuring cell with metal-sealed feed through
  • with universal power supply-input voltage from 90 ~ 264VAC, frequency 47 ~ 440Hz
  • integrated set point adjustment and relay output for both vacuum and temperature
  • Easy To Read 4 Digit Backlit Display
  • Change Cable Length Without Re-calibrating upto 8 meters
  • S 485 computer interface with data logging - optional
  • Technical specifications:

    MODLES DVTC-001
    Number of sensor channels ONE EACH FOR VACUUM AND TEMPERATURE
    Display Range FULL range: 0.001 to 1000 mbar for vacuum & -100 to +800 Deg C for temp.
    Measuring Range USEFUL range: 0.001mbar to 100 mbar and -100 to +800 Deg C for temp.
    Units of Display m Bar and Deg C
    Vacuum Gauge Pirani (S-PR-GH sensor tube) and RTD thermo couple
    Temp. Range +2 to +50 deg. Celsius
    Display 4-digit bright red LED, 14 mm high
    Display Indicators   red LED for SET PT
    VAC and ATM Adjust Adjusted by trim pot at the signal conditioner PCB.
    Process control set points ONE EACH FOR VAC & TEMP
    Process Control Relays Two relays, 2 amp, 240 VAC contacts; independent available through the DB15 accessory connector
    Nonvolatile Memory for SET POINTS
    Operating Voltage universal power supply, which operates on input voltages from 90 ~ 264VAC 50/60 Hz; standard IEC 320 instrument power input receptacle on rear panel; replacement fuse type: 5 mm X 20 mm, 2 amp
    Mounting Side clips are provided for panel mounting
    Environmental Considerations not for use with explosive or corrosive gases
    Overall dimensions 200W X 210D X110 H (mm)